TECHNICAL SPECIFICATIONS
Recommended Applications
Neonatal, Infant, and Pediatric Ventilation
Flow
0-32 LPM, Flush 40 LPM
FiO2
0.21 to 1.00
Mode
OFF, CPAP, VENT
Inspiratory Time
0.10 to 2.90 sec.
Expiratory Time
0.30 to 60.0 sec.
I : E Ratio
10:1 to 1:600 (Readout 1:0.1 to 1:99)
Rate
1 to 150 BPM
Expiratory Pressure
2 to 20 cm H2O; Zero Baseline possible at 24 LPM and higher rates
Inspiratory Pressure
5 to 70 cm H2O
Manual Breath
Yes
Alarms
Independent for Microprocessor; Low Airway Pressure, Leaks, Patient Disconnect, Fail to Cycle, Apnea, Prolonged Inspiration, and Peak Overpressure
Alarm Delay Time
3 to 60 sec.
Alam Mute
30 sec. ± 5 sec. |
Safety Pressure Relief Valve
Yes (adjustable)
Indicators
Inspiratory Time, Expiratory Time, I:E Ratio, Rate, Inspiratory Phase, Inverse I:E Ratio, Mode Selection, Alarm, High and Low Alarm Set Points, and Mean Airway Pressure
Manual Teats
Checks microprocessor function, displays, alarm function
Safety Auto Lock-Out Circuit
Yes, Inspiratory Phase terminated at 4 sec. Even if microprocessor fails
Timing Pre-set
Yes
Dimensions
Height: 13" (33.02 cm)
Width: 14" (35.56 cm)
Depth: 9" (23.00 cm)
Weight
Net: 21 lbs. (9.54 Kg)
Shipping
30 lbs. (13.63 Kg)
Power Requirements
117 VAC 50/60 Hz, Circuit Breaker Protected, 3-Wire SJT Power Cord Provided, Hospital-Grade Plug, Current leakage less than 20 micramps when tested per UL 544
Transport Power Requirements
12 VDC 1.2 amps |