Sechrist Products ›› Respiratory Therapy Ventilators › Sechrist IV-100B Infant Ventilator

The Sechrist IV-100B Infant Ventilator addresses infant ventilation needs with easy-to-use controls that permit rapid setting of ventilation parameters and allow clinicians to fine tune to the greatest accuracy.
The built-in displays provide easy-to-read continuous ventilation parameter information. The alarm system is comprehensive, but avoids information overload during alarm conditions. To prevent inadvertent PEEP when higher flow rates are required, a special exhalation value block and pediatric circuit are available.
The IV-100B is capable of meeting the most demanding requirements of the intensive care unit. You can ventilate newborn to pediatric patients with precise control of oxygen concentration, at respiratory rates of 1 to 150 breaths per minute. The Inspiratory and Expiratory Pressure controls are non-interactive. A simple, yet comprehensive alarm system monitors the parameters you have set up and notifies you of significant changes.

 
Brochure
Exhalation Valves and Circuits
Humidifiers and Accessories
Monitors and Accessories
Ventilator Accessories and Parts

TECHNICAL SPECIFICATIONS

Recommended Applications
Neonatal, Infant, and Pediatric Ventilation

Flow
0-32 LPM, Flush 40 LPM

FiO2
0.21 to 1.00

Mode
OFF, CPAP, VENT

Inspiratory Time
0.10 to 2.90 sec.

Expiratory Time
0.30 to 60.0 sec.

I : E Ratio
10:1 to 1:600 (Readout 1:0.1 to 1:99)

Rate
1 to 150 BPM

Expiratory Pressure
2 to 20 cm H2O; Zero Baseline possible at 24 LPM and higher rates

Inspiratory Pressure
5 to 70 cm H2O

Manual Breath
Yes

Alarms
Independent for Microprocessor; Low Airway Pressure, Leaks, Patient Disconnect, Fail to Cycle, Apnea, Prolonged Inspiration, and Peak Overpressure

Alarm Delay Time
3 to 60 sec.

Alam Mute
30 sec. ± 5 sec.

 

Safety Pressure Relief Valve
Yes (adjustable)

Indicators
Inspiratory Time, Expiratory Time, I:E Ratio, Rate, Inspiratory Phase, Inverse I:E Ratio, Mode Selection, Alarm, High and Low Alarm Set Points, and Mean Airway Pressure

Manual Teats
Checks microprocessor function, displays, alarm function

Safety Auto Lock-Out Circuit
Yes, Inspiratory Phase terminated at 4 sec. Even if microprocessor fails

Timing Pre-set
Yes

Dimensions
Height: 13" (33.02 cm)
Width: 14" (35.56 cm)
Depth: 9" (23.00 cm)

Weight
Net: 21 lbs. (9.54 Kg)

Shipping
30 lbs. (13.63 Kg)

Power Requirements
117 VAC 50/60 Hz, Circuit Breaker Protected, 3-Wire SJT Power Cord Provided, Hospital-Grade Plug, Current leakage less than 20 micramps when tested per UL 544

Transport Power Requirements
12 VDC 1.2 amps