ERKENDE REINIGERS
It is the user’s responsibility to determine the appropriateness of any products prior to actual use and to be cognizant that reformulation of these products by the manufacturer may occur without notice. Compatibility with acrylic, seals, metallic components, paints, and oxygen should all be considered.
CARE AND DISINFECTION OF THE CHAMBER AND GURNEY
-
Clean the chamber according to the type of cases being treated and as directed by the medical staff. Refer to the list of acceptable cleaners available from Sechrist Industries.
-
Wash Stretcher, Pillows. Positioning Pads and Mattress surfaces with approved cleaner or mild dishwashing soap. Refer to the Gurney User Manual for cleaning instructions.
CLEANING OF PAINTED AND ANODIZED SURFACES
-
Use the same methods for cleaning as used for the Acrylic Cylinder.
ACCEPTABLE BIOCIDES per ASME PVHO-2 MOST RECENT EDITION
-
Bleach, up to 15% aqueous sodium hypochlorite.
-
Aqueous hydrogen peroxide, 3% to 20%.
-
Aqueous chlorine dioxide, up to 2%.